Calculation of long-wave radiation

ISFS has been using a combination of Campbell Scientific NR01 4-component radiometers and Epply pyrgeometers to measure long-wave (infrared) radiation. These radiometers measure the response of a thermopile within the radiometer dome (Rpile), which responds to the difference in radiation from the field of view minus the equivalent black-body radiation of the sensor housing (measured by Tcase). Thus, to first order the output of this sensor should be

Rlw = Rpile + SB * Tcase^4



where Rlw is the longwave radiation and SB is the Stefan-Boltzmann constant,

5.67e-08 W/m^2/degK^4

However, since the transmissivity of the radiometer's dome is not exactly 1 in the infrared, the sensor is slightly sensitive to the dome's temperature (Tdome). Also, the emissivity of the dome is not exactly 1 in the visible, so the sensor also is slightly sensitive to incoming short-wave radiation (Rsw). Thus, to second order, the output of this sensor should be

Rlw = Rpile + SB * ( Tcase^4 + B * ( Tcase^4 - Tdome^4 ) ) + f * Rsw



where B and f are sensor-specific values determined by shading tests performed in clear-sky conditions. In Colorado, we generally are able to achieve a change in Rsw of 600-800 W/m^2 to determine the change in Rlw (and thus f). B is initially determined in the laboratory by measuring the response to changing Tdome and Tcase, but we found it necessary to adjust these values to obtain a "square-wave" response during the shading tests. Values of B for our sensors range from 1.8-3.0 and of f from 0.5-1.9%.

Obviously, Rsw is either "in" or "out" corresponding to whether Rlw is "in" or "out".

We note that standard Epply pyrgeometers have one Tdome sensor (a thermistor attached to the inside of the dome), but several studies have shown that temperature gradients across the dome can result in significant errors. All of ISFS's Epply pyrgeometers have 3 thermistors to measure Tdome (at least since 1998), though we wire them in series and record only the average of all 3. (This is slightly in error, since the (non-linear) calibrations of the 3 thermistors could differ.)

Finally, we note that we have changed the use of Kipp and Zonen pyrgeometers (CG4), which have a shallower dome that is in close thermal contact with the case. Since Tdome-Tcase is significantly smaller for these sensors, a correction using B is not needed.