Post-T-TRH
Post calibration check for CME04

Temperature sensors checked over the range of -10C to 40C

 

Tower:height Sensor mean error
    (Sensor - Tref)
Willow:17m TRH007 0.038
Willow:10m TRH001 0.041
Willow:6m TRH204 -0.016
Willow:2m TRH008 0.053
Willow:1m TRH703 -0.047
Pine:30m TRH704 0.022
Pine:10m TRH101 0.019
Pine:6m TRH104 0.050
Pine:2m TRH701 0.033
Pine:1m TRH004 0.004
Aspen:30m TRH005 0.030
Aspen:10m TRH006 0.022
Aspen:6m TRH003 -0.100
Aspen:2m TRH702 0.017
Aspen:1m TRH502 0.015