HDP Notes 06-04-19 and 20 Installed AIRCOA E with two inlet lines. Fixed positive pressure leak (0.2 kPa/min), but it returned several hours after leaving site. 06-07-06 S. Schaeffer visits to fix high pressure leak check. Replaces broken fitting but leak still present. Leaves unit on, but communications appears to have failed. 06-07-12 S. Schaeffer returns, replaces Bev-a-line to micropump. Leak rate drops to ~0.1 kPa/min. Communication restored. 06-08-22 00:21 changed settings to purge nafion with line 2 during low pressure leak checks instead of line 1 06-09-25 High pressure leak rate suddenly jumps from -0.09 kPa/min to -2.2 kPa/min. No known visitors. 06-10-11 Updated code to new version (lll) in anticipation of switching flow meter wires. RH jumped ~ 0.65. 06-10-11 HDP site visit. Replaced "Air" solenoid to fix leak. Zero and span on LI820. Tuned up. 06-12-24 Low P leak check suddenly jumps to -0.25 kPa/min. 07-02-04 G. Poulous onsite. Breath test and outer line connections tightened. 07-03-07 B. Stephens on site. Replaced 4 calibration cylinders. Replaced air select solenoid (6) which fixed a leak. Switched lines 2 and 3 at bulkheads to diagnose offset. Fixed settings2.txt to go to line 2 instead of line 1 during cals. 07-07-10 Micropump failed. Stopped program on 7/11 07-07-18 D. Bowling replaced pump, tuned up. Humitter appears to have intermittent connection problem. 07-12-16 Memory stick filled up. Cleared and rebooted on 07-12-18. Somehow pen drive did not mount. Manually mounted and restarted program on 07-12-19. 08-01-25 Both HS2 and LS2 appear to have run out. Stopped program on 08-01-30. 08-02-23/24 Replaced 4 calibration cylinders. Replaced Li-820 (new = S/N CGA-689), mole-seives, and humitter. Replaced line 2 and 3 5 um line filters and rebuilt inlet manifold to put filters upstream of needle-valves. Replaced inlet filters and repositioned 17.7 m inlet. Swapped lines 2 and 3 back to original configuration (line 3 = 17.7 m west inlet). 08-07-15 Found NTP was not working. 68 Sec slow. Set manually 08-08-28 NTP still not working. 5:05 slow. Set manually at 20:43:30 08-09-01 Appears to have ingested liquid water during intense storm ~ 5:30 AM UTC. Some components recovered, but flow low, V6 leaks, and Li820 cal off. Stopped program and commented out line in rc.sysinit 08-09-08.