Project #2002-111 IDEAS Phase 1

Instrument Development and Education in Airborne Science

J. Stith, et al.

NSF/NCAR C-130Q Hercules (N130AR)


Aerosol Racks - ADS Sample Table

Dave Rogers - 3/20/02

NOTES: 1. RDMA setup - same as EPIC
2. Nephs serial is free running, about 2Hz. How to sample?
3. Climet serial output looks weird.
4. Inlet ident - can we rig up and log a 3-position switch ?
5. CN counter #1 = same as always. CN #2 has flow but no temperature or pressure.

Forward Starboard Rack
instrument variable description units type rate comments
RDMA
(same as EPIC)
RDMAHVPS High Voltage Power Supply v RS-232 1 Hz unpack 10 variables from string
  RDMAPABS Absolute Pressure atm RS-232 1 Hz  
  RDMAQA Aerosol Inflow v RS-232 1 Hz  
  RDMAQEX Excess Outflow v RS-232 1 Hz  
  RDMAQS Classified Aerosol Outflow v RS-232 1 Hz  
  RDMAQSH Sheath Inflow v RS-232 1 Hz  
  RDMARH Relative Humidity % RS-232 1 Hz  
  RDMATEMP Sample Temperature C RS-232 1 Hz  
  RDMAVD Dilution Valve Control Volts v RS-232 1 Hz  
  RDMAVSH Sheath Valve Control Volts v RS-232 1 Hz  
  (string) sizedistup*.txt cm^-3 RS-232   "long string, once every 156 sec"
  (string) sizedistdn*.txt cm^-3 RS-232   "long string, once every 156 sec"
  (string) "ambient*.txt (flow, press, T, rh)"   RS-232   "long string, once every 156 sec"
  (string) countrec*.txt   RS-232   "long string, once every 156 sec"
neph-1 xnbscat1 back scatter 1/m RS-232 1 Hz ? unpack 5 variables from string
  xncal1 calib coeff 1/m RS-232 1 Hz ?  
  xnp1 pressure mb RS-232 1 Hz ?  
  xnt1 temperature K RS-232 1 Hz ?  
  xnrh1 relative humidity % RS-232 1 Hz ?  
neph-2 xnbscat2 back scatter 1/m RS-232 1 Hz ? unpack 5 variables from string
  xncal2 calib coeff 1/m RS-232 1 Hz ?  
  xnp2 pressure mb RS-232 1 Hz ?  
  xnt2 temperature K RS-232 1 Hz ?  
  xnrh2 relative humidity % RS-232 1 Hz ?  
neph-humidifier xnht saturator temperature C analog 1 Hz does not exist yet
Climet xconclim [6] "particle conc array, six size bins" #/vol RS-232 1 Hz unpack 8 variables from string
  xtclim temperature K RS-232 1 Hz  
  xrhclim relative humidity % RS-232 1 Hz  
Wyoming heated inlet xwyof flow rate LPM analog 1 Hz  
  xwyot temperature of inlet tip C analog 1 Hz  
Inlet select xinlet switch (thumbwheel?) inlet selected
(three-position switch if possible)
  ? 1 Hz  

Aft Starboard Rack
instrument variable description units type rate comments
CN - 1 CONCN Condensation Nuclei (CN) Concentration   calc 10 Hz  
  CNTEMP CN Counter Inlet Temperature   analog 10 Hz  
  CNTS TSI CN Counter Output   TTL pulses 10 Hz  
  PCN CN Counter Inlet Pressure   analog 10 Hz  
  FCN Raw CN Counter Sample Flow Rate   analog 10 Hz  
  FCNC Corrected CN Counter Sample Flow Rate   calc 10 Hz  
CN-2 CONCN2 Condensation Nuclei (CN) Concentration   calc 10 Hz  
  CNTS2 TSI CN Counter Output   TTL pulses 10 Hz  
  FCN2 Raw CN Counter Sample Flow Rate   analog 10 Hz  
  FCNC2 Corrected CN Counter Sample Flow Rate   calc 10 Hz  

Right Instrument Pod
instrument variable description units type rate comments
DMT - total water xtwc total water content g/m^3 analog 10 Hz  

Left Instrument Pod
instrument variable description units type rate comments
Nevzorov xntwc total water content g/m^3 analog 10 Hz  
  xnLwc liquid water content g/m^3 analog 10 Hz  
Maybe additional variables for Nevzorov ..?